Remedi Portable Dental X-Ray Device
Regular price
$3,000.00
Couldn't load pickup availability
You can contact us for your orders and questions. +90543 456 2614
All your orders are sent by fast air cargo with UPS or DHL Shipping
Your orders will be prepared and shipped within 3 business days.
Customs clearance in your country is not the seller's responsibility.
REMEX-T (K) 100 Portable X-Ray Equipment
The REMEX-T (K) 100 Portable X-ray Equipment is designed for use by trained dentists and dental technicians as an extra-oral x-ray source to produce diagnostic x-ray images using intra-oral imaging. It is intended for use on adults, children, and pregnant women.
Responsibilities and Compliance:
- The owner/operator is responsible for verifying ongoing compliance with exposure limits, leakage radiation, beam alignment, and calibration of kVp and mA.
- Annual verification may be required by a qualified service technician, as mandated by federal law.
- Compliance with local, state, and federal regulations is the responsibility of the owner/operator.
Cleaning Instructions:
- Ensure the adapter is unplugged before cleaning.
- Ensure the power is off on the REMEX-T (K) 100 during cleaning.
- Use only an alcohol-based disinfectant with a tissue or cloth moistened with liquid or spray.
- The REMEX-T (K) 100 and its adapter must not be subjected to any sterilization procedure.
Specifications:
- Classification: Class IIb (Annex IX, Rule 10, Council Directive 93/42/EEC as amended by Directive 2007/47/EC)
- Model: REMEX-T100, REMEX-K100 (model name varies by X-ray tube)
-
Electric Shock Protection Type:
- Class II equipment (Exposure mode)
- Built-in power supply equipment (Charge mode)
- Type B Applied part
Power Specifications:
-
AC/DC Adapter:
- Input: 100-240 Vac, 50/60 Hz, 1.5 A
- Output: 12.6 Vdc, 1.5 A
-
Rechargeable Battery:
- 11.1 Vdc, 1500 mAh
- Power Input: 160 VA (in charging mode)
X-Ray Tube Specifications:
- Tube Voltage: 70 kV (Fixed)
- Tube Current: 2 mA (Constant)
- Exposure Time Range: 0.01 s ~ 1.3 s
- Focal Spot Size: 0.4 mm (conforming to IEC 60336: 1993)
- Natural Filtration: Min. 1.0 mmAl / 0.8 mmAl
Filament Characteristics:
- 1.0 ~ 4.0 V, 2.2 ~ 3.0 A (max filament current)
- 2.0 ~ 3.5 V, 2.2 ~ 3.0 A (max filament current)
Anode Specifications:
- Angle: 12.5°/12.0°
- Thermal Properties: 4.3 kJ / 7.0 kJ
- Maximum Anode Heat Dissipation Rate: 430 W / 560 W
Protection and Operation:
- Water and Particulate Ingress Protection: IPX0
-
Mode of Operation: Continuous
- Recharge time of the high voltage tank is 10 sec.
- Expected Service Life: 5 years
Basic Performance:
-
Accuracy of Loading Factors:
- Tube voltage accuracy: less than 10%
- Tube current accuracy: less than 20%
- Irradiation time accuracy: less than 5% + 50 ms
- Tube current time accuracy: ± (10% +0.2 mAs)
-
Repeatability of Radiation Output:
- Coefficient of variation of measured air kerma values: less than 0.05